SAMM 592 Scope of Accreditation

Scope of Accreditation – SAMM No. 592

Accredited to MS ISO/IEC 17025:2017 by Standards Malaysia under the Laboratory Accreditation Scheme of Malaysia (SAMM).
Fields of calibration: Pressure, Force, Mass, Temperature & Dimensional.
Schedule issue date: 14 January 2025  ·  Valid until: 16 April 2030

Pressure – Permanent laboratory

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Pressure measuring device – Hydraulic100 psi to 300 psi0.06 psiCalibrate using dead weight tester as standards with reference to DKD-R 6-1 sequence A, C
300 psi to 1000 psi0.2 psi
1000 psi to 3000 psi0.6 psi
3000 psi to 5000 psi0.9 psi
5000 psi to 10,000 psi2 psi
Pressure measuring device – Pneumatic0 bar to 2 bar0.001 barCalibration using digital pressure test gauges as standards with reference to DKD-R 6-1 sequence A, C
2 bar to 20 bar0.01 bar
20 bar to 70 bar0.03 bar
Pressure measuring device – Vacuum-1 bar to 0 bar0.005 barAs above
Pressure measuring device – Hydraulic0 psi to 1000 psi0.4 psiAs above
1000 psi to 5000 psi2.4 psi
5000 psi to 10,000 psi4 psi

Pressure – On site

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Pressure measuring device – Pneumatic0 bar to 2 bar0.001 barCalibration using digital pressure test gauges as standards with reference to DKD-R 6-1 sequence A, C
2 bar to 20 bar0.01 bar
20 bar to 70 bar0.03 bar
Pressure measuring device – Vacuum-1 bar to 0 bar0.005 barAs above
Pressure measuring device – Hydraulic0 psi to 1000 psi0.4 psiAs above
1000 psi to 5000 psi2.4 psi
5000 psi to 10,000 psi4 psi

Force – Permanent laboratory

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Push pull gauge / digital force gauge / tension gauge0 kgf to 0.5 kgf0.001 kgfCalibrate using standard weight and poise weights. Calibrations may be given in other units by conversion from SI units.
0.5 kgf to 1 kgf0.002 kgf
1 kgf to 3 kgf0.005 kgf
3 kgf to 50 kgf0.01 kgf
50 kgf to 100 kgf0.1 kgf

Mass – Permanent laboratory

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Standard weight1 g0.04 mgCalibrate using reference standard weight by comparison method according to ABBA weighing scheme
2 g0.05 mg
5 g0.06 mg
10 g0.07 mg
20 g0.09 mg
50 g0.10 mg
100 g0.17 mg
200 g0.4 mg
500 g0.002 g
1 kg0.006 g
2 kg0.02 g
5 kg0.03 g
10 kg0.2 g
20 kg0.4 g

Mass – On site

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Electronic balanceUp to 50 g0.0002 gCalibrate using standard weights as standards based on ASTM E898-20
Up to 100 g0.0004 g
Up to 200 g0.001 g
Up to 600 g0.002 g
Up to 1000 g0.01 g
Up to 2000 g0.02 g
Up to 4000 g0.04 g
Up to 10000 g0.1 g
Up to 25000 g0.4 g
Up to 30000 g0.5 g
Up to 60 kg0.02 kg
Up to 100 kg0.04 kg
Up to 300 kg0.05 kg
Up to 500 kg0.1 kg
Up to 700 kg0.2 kg
Up to 1200 kg0.4 kg
Up to 1500 kg0.5 kg

Temperature – Permanent laboratory

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Temperature indicating instrumentType K: -100 °C to 1300 °C1.0 °CCalibrated by electrical simulation using temperature calibrator
Type J: -100 °C to 1100 °C0.9 °C
Type E: -100 °C to 990 °C0.9 °C
Type T: -100 °C to 390 °C1.0 °C
Type S: 0 °C to 1700 °C1.4 °C
Type R: 0 °C to 1700 °C1.3 °C
Pt100: -100 °C to 800 °C0.3 °C
Temperature sensor with display unit-30 °C to 140 °C0.3 °CCalibrate by comparison method using PT100 as reference standard, in liquid bath and temperature block calibrator
141 °C to 600 °C0.7 °C
Thermohygrograph / thermohygrometer and humidity related equipment30 %RH to 90 %RH3.5 %RHCalibrate by comparison method using thermohygrometer in humidity chamber
20 °C to 35 °C0.7 °C

Temperature – On site

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Temperature indicating instrumentType K: -100 °C to 1300 °C1.0 °CCalibrated by electrical simulation using temperature calibrator
Type J: -100 °C to 1100 °C0.9 °C
Type E: -100 °C to 990 °C0.9 °C
Type T: -100 °C to 390 °C1.0 °C
Type S: 0 °C to 1700 °C1.4 °C
Type R: 0 °C to 1700 °C1.3 °C
Pt100: -100 °C to 800 °C0.3 °C
Temperature sensor with display unit-30 °C to 140 °C0.3 °CCalibrate by comparison method using PT100 as reference standard, in liquid bath and temperature block calibrator
141 °C to 600 °C0.7 °C

Dimensional – Permanent laboratory

Instrument calibrated / measurement parameterRangeCMC expressed as uncertainty (±)*Remarks
Caliper – external measurement0 mm to 300 mm0.01 mmCalibrate using gauge blocks as standards with reference to JIS B 7507:2016
300 mm to 600 mm0.02 mm
Caliper – internal measurement0 mm to 300 mm0.01 mmAs above
External micrometer25 mm travel range1 µmCalibrate using gauge blocks as standards with reference to JIS B 7502:2016
Frame size up to 100 mm1 µm
Frame size 100 mm to 225 mm2 µm
Frame size 225 mm to 425 mm3 µm
Frame size 425 mm to 500 mm4 µm

* The uncertainty covered by the CMC is expressed as the expanded uncertainty corresponding to a coverage probability of approximately 95% with a coverage factor of k=2, unless stated otherwise. This page is a summary; the official schedule (PDF) prevails.